Yamato-scientific WR600S AutoPure User Manual

Browse online or download User Manual for Equipment Yamato-scientific WR600S AutoPure. Yamato Scientific WR600S AutoPure User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Autopure

Operating and Maintenance manual Autopure WR600A WR600G WR600B WR600S Yamato Scientific co., ltd

Page 2 - INTRODUCTION

OPERATING THE AUTOPURE WR600 SYSTEM Modes of operation Your Autopure WR600 system has a number of operating modes which can be activated via the

Page 3 - System Schematic

Additional displays System maintenance messages DISPLAYED MESSAGE ACTION SYSTEM STATUS EXCH. CARTRIDGES Occurs automatically. See MAINTENAN

Page 4 - Characteristics

Alarm displays DISPLAYS STATUS of the SYSTEM CARTRIDGE OUT The Autopure WR600 has stopped operating because either the QUANTUM or Q-Gard pu

Page 5 - System performances

Use of the Keypad The keypad allows the user to activate the different operating modes or to review information about system performance. KEYPAD

Page 6 - INSTALLATION

Use of Keypad continued KEYPAD ACTION DISPLAY MENU function continued Printer In either PRE-OPERATE or PRODUCT Press MENU for at least 2 sec

Page 7 - Installation

Use of Keypad continued KEYPAD ACTION DISPLAY Cleaning the A10 module (1 hour duration) The A-10 CLEANING mode is used to clean the oxidation ch

Page 8 - S R . N ° Z 5 A L 5 6 8 A 5 C

Using the keypad continued KEYPAD ACTION DISPLAY MENU function continued Non temperature compensated resistivity display. In either PRE OPERATE

Page 9

Servicing the Autopure WR600 The intermittent recirculation of water inside the Autopure WR600 allows the cartridge(s) to become fully hydrated

Page 10 - Routinely seen displays

MAINTENANCE Table of periodic maintenance Every month Annually (once a year) Following a message on the LCD Note the operating parameters for

Page 11 - Additional displays

This is a short sanitisation maintenance program. This allows the UF module to be sanitised at the start of an evening and throughout that night. C

Page 12

INTRODUCTION General Information The Autopure WR600 system is used as a final step of water purification. The feedwater to a Autopure WR600 syst

Page 13 - Use of the Keypad

Replacing the Q-Gard pack (figure 4) 1. Put the Autopure WR600 system into STANDBY mode by pressing OPERATE / STANDBY for 2 seconds. Do not turn

Page 14 - Use of Keypad continued

Cleaning the A10 After performing several A-10 TOC measurements, the detection cell in the A-10 needs to be cleaned of any built- up residual organ

Page 15

MAINTENANCE Maintenance messages Whenever the SERVICE LED is blinking, a message is displayed on the screen which indicates the nature of the servic

Page 16 - Using the keypad continued

STATUS / PROBLEM CAUSE REMEDY START SANIT. A cleaning cycle of the UF cartridge needs to be started. Start a cleaning procedure. See the MAINTE

Page 17

List of ALARM errors and ALARM messages When the ALARM LED is flashing, an error code number will be displayed which indicates the nature of the pr

Page 18 - MAINTENANCE

A-10 ERROR # DESCRIPTION REMEDY 0 EEPROM savings error If the error persists, then contact Yamato Technical Service. 1 Error in analogue to

Page 19 - KEYPAD ACTION DISPLAY

APPENDIX 1 Interrupting the sanitisation cycle of an UF cartridge If a sanitant chemical has been introduced into the Autopure WR600 , it is absol

Page 20

Replacing the main electrical power fuse 1. Put the system into STANDBY mode by pressing OPERATE / STANDBY for 2 seconds. 2. Turn off the pow

Page 21

Warranty Millipore Corporation (“Millipore”) warrants the products manufactured by it against defects in materials and workmanship when used in acc

Page 22

System Schematic The water flow schematic of a Autopure WR600 system is shown below. Only the main components are shown.

Page 23

Characteristics Composition of materials in contact with water Pieces Material Pieces Material Pack Adapter : Q-Gard Pack Inlet solenoid valve : Q

Page 24

Environmental Conditions Ambient storage and operating temperature Humidity 5°C < T < 40°C 20 % - 80 % without condensation Hydrau

Page 25

INSTALLATION Unpacking The different components supplied with the system are shown in figure 2 and are listed below. (A) Autopure WR600 wa

Page 26 - APPENDIX 1

Installation Connection of feed water to the system (Figure 3) System fed from a reservoir : 1. Cut the feed water tubing, 8 mm OD (Figure 2,

Page 27

Installation of the Q-Gard purification pack (Figure 4) (The Q-Gard is only used on Autopure WR600 systems equipped with a pack adapter) 1. Raise

Page 28 - Warranty

Rinsing the Autopure WR600 for the first time A 5 minute rinsing cycle “AIR PURGE” occurs automatically whenever new cartridges are installed in t