Yamato-scientific BM210 Water Baths User Manual

Browse online or download User Manual for Equipment Yamato-scientific BM210 Water Baths. Yamato Scientific BM210 Water Baths User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 16
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
WATER BATH
BM110
BM210
First Edition
Thank you for your Yamato Scientific BM series Water Bath purchase.
For proper use of this unit, please read the instruction manual and warranty thoroughly before
operation. Keep both for any future references.
Yamato Scientific
Read and apprehend important warnings in this
instruction manual prior to use.
Page view 0
1 2 3 4 5 6 ... 15 16

Summary of Contents

Page 1 - WATER BATH

WATER BATHBM110BM210First Edition● Thank you for your Yamato Scientific BM series Water Bath purchase.● For proper use of this unit, please read the

Page 2 - Table of Contents

8Identification of PartsSample : Combined with RE210Combination with RE210According to the position or space to set the bath and the body, the arm jac

Page 3 - 1. Specifications

9Identification of PartsSample : Combined with RE510Combination with RE510

Page 4 - 2. Safety Information

10 4. How to OperateHow to operate1. Pour adequate water (3000ml for BM110, 5000ml for BM210) into the tank, and place a thermometerin the holder. Us

Page 5 - Safety Information

115. MaintenanceDaily maintenance/Weekly maintenanceDo not disjoint the unitNever modify the unit. There are some high-voltage areas inside the unit,

Page 6 - Safety Information

126. After Sale Service and WarrantyRequest for RepairWhen you request repairIf any troubles should occur, stop the operation immediately, turn the po

Page 7

13 7. Replacement PartsName of Parts Parts Number SpecificationFuse 2-10-006-0002Φ6.4×30 250V 4A (for BM110)Fuse 2-10-006-0003Φ6.4×30 250V 7A (fo

Page 8 - 3. Identification of Parts

14 8. Wiring DiagramBM110BM210Symbol Name of Parts Symbol Name of PartsF Fuse S Service OutletH Heater SW Power SwitchTF Thermal Fuse TCS Thermoregul

Page 9

Table of Contents1. Specifications...

Page 10 - Identification of Parts

11. SpecificationsModel BM110 BM210Temp. range during operationRoom temp. +5℃ ~ 95℃ *1Temp. setting range0℃ ~ 100℃Accuracy of temp. adjustment *2±2℃ a

Page 11

22. Safety InformationSafety SymbolsSafety InformationThis instruction manual and our products apply various indications for safety. Ignoring theseind

Page 12 - 4. How to Operate

3 Safety InformationSafety PrecautionsDo not use this unit for any purpose other than its intended use, described in this manual.Do not use this unit

Page 13 - 5. Maintenance

4Safety InformationSafety PrecautionsIf it begins to thunder…In the event of electrical storm turn off the main power. Neglecting this procedure can r

Page 14 - Request for Repair

5Safety InformationHazardous MaterialsHazardous material are listed below. Never use these materials as samples orheat media.Nitroglycol, Nitroglyceri

Page 15 - 7. Replacement Parts

63. Identification of PartsBM110BM110Temperature DialPower CablePower SwitchFusePower LampSensorHeaterThermometerThermometer Holder

Page 16 - 8. Wiring Diagram

7Identification of PartsBM210BM210Temperature DialPower SwitchDrain PlugPower CableFusePower LampHeaterSensorThermometer HolderThermometer